To begin, add two milliliters of MIRO five buffer to the Oroboros chamber. Cover the chamber with a stopper without introducing air bubbles. Then carefully pull the stopper to form a single air bubble.
Open the Oroboros DatLab software in the block temperature field. Enter a temperature of 25 degrees Celsius and click on Connect to Oxygraph-2k. Click and select the folder where the calibration is to be saved.
And click on save. When a new window opens, name the experiment and the samples. Then click on save.
Next, go to the layout menu and choose the first option one, calibration experiment. Graph three, temp. When the red line shows a standard straight line with points around zero picomole, select two points, one for chamber A and the other for chamber B.Select oxygen concentration on the right of the screen.
Select the marked point and copy the temperature and barometric pressure values relative to the point. Paste these values in the boxes below, and click on calibrate and copy to clipboard. Then go to the file menu and select save and disconnect.