Begin by preparing the mobile phases for high-performance liquid chromatography, or HPLC. For mobile phase A, or MPA, use five millimolar ammonium formate in water with 0.05%formic acid. For mobile phase B, or MPB, use 0.05%formic acid in a one-to-one methanol acetonitrile mixture.
To begin the HPLC method development, open the liquid chromatography or LC software and create a new protocol. Then, click the Edit Method button to edit the binary gradient. Under the Binary Gradient tab, set the timestamps and concentrations of MPA and MPB.
Set the stop time to 18 minutes, the flow rate to 0.400 milliliters per minute, and the minimum and maximum pressure limits to zero and 6, 000 PSI respectively. To initialize the HPLC, insert the LC column and column guards in the HPLC column section. Carefully observe the minimum and maximum column pressures.