Accedi

Accurate calibration of glassware, such as volumetric flasks, pipettes, and burettes, is essential to ensure accurate measurements in the analytical laboratory. Calibration helps maintain consistency across measurements and prevents errors arising from inaccurate volumes.

Volumetric flasks:Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled water until the meniscus (the curved surface of the liquid) aligns with the calibration mark. Viewing the meniscus at eye level ensures accurate readings. Calibration is typically performed at a specific temperature (~20°C) to account for the effect of temperature on volume. If the water level is above the line, drain it carefully until it hits the mark; if the water line is below, add more water dropwise until it reaches the mark. After calibration, the flask's capacity can be confirmed by weighing the water and comparing it with the theoretical value based on its density.

Pipettes:Pipettes are used to transfer specific volumes of liquid.

Volumetric Pipettes: These are designed to deliver a single, highly accurate volume, typically to four or five significant figures. For calibration, the pipette is first cleaned and rinsed with the solution to be measured, filled to the calibration mark, and allowed to drain completely into a container. The accuracy is then checked by weighing the water delivered and comparing this with the theoretical volume based on water density.

Graduated Pipettes: Unlike volumetric pipettes, graduated pipettes are marked with a series of lines and can deliver multiple volumes but with slightly lower accuracy—generally up to three significant figures. The meniscus must also be carefully observed during calibration and measurement.

Burette:A burette is a long, graduated glass tube with a tap at one end that accurately dispenses variable amounts of a solution, particularly in titrations. When calibrating a burette, ensure that the meniscus is correctly aligned with the markings before and after titration, and check that the volume dispensed is as expected by comparing the weight of the water delivered to its theoretical value. The burette's reservoir contains markings to indicate the volume of the liquid. The stopcock or tap at the bottom serves as a manual valve that allows users to control the liquid flow from the reservoir, allowing for precise measurement of the dispensed amount.

Other Glassware:Other types of laboratory glassware, such as graduated cylinders, are also commonly used for volume measurements but offer lower accuracy than volumetric equipment. They typically deliver volumes up to three significant figures, suitable for less precise work. However, calibration of these items is still important for maintaining measurement reliability.

Calibrating laboratory glassware ensures accurate and reliable measurements, which is crucial for maintaining the integrity of analytical results. Whether using volumetric flasks, pipettes, burettes, or other glassware, proper calibration techniques help prevent errors and improve the precision of laboratory work.

Dal capitolo 9:

article

Now Playing

9.8 : Glassware Calibration

Method Development and Sampling Techniques

141 Visualizzazioni

article

9.1 : Development of Analytical Methods

Method Development and Sampling Techniques

333 Visualizzazioni

article

9.2 : Quality Control

Method Development and Sampling Techniques

122 Visualizzazioni

article

9.3 : Quality Assurance

Method Development and Sampling Techniques

95 Visualizzazioni

article

9.4 : Data Validation

Method Development and Sampling Techniques

117 Visualizzazioni

article

9.5 : Qualitative Analysis

Method Development and Sampling Techniques

176 Visualizzazioni

article

9.6 : Quantitative Analysis

Method Development and Sampling Techniques

207 Visualizzazioni

article

9.7 : Instrument Calibration

Method Development and Sampling Techniques

133 Visualizzazioni

article

9.9 : Standard Solutions

Method Development and Sampling Techniques

160 Visualizzazioni

article

9.10 : Blank Solutions

Method Development and Sampling Techniques

179 Visualizzazioni

article

9.11 : Sampling Methods: Overview

Method Development and Sampling Techniques

212 Visualizzazioni

article

9.12 : Sampling Methods: Sample Types

Method Development and Sampling Techniques

153 Visualizzazioni

article

9.13 : Sampling Plans

Method Development and Sampling Techniques

154 Visualizzazioni

article

9.14 : Sample Preparation for Analysis: Overview

Method Development and Sampling Techniques

155 Visualizzazioni

article

9.15 : Sample Preparation for Analysis: Advanced Techniques

Method Development and Sampling Techniques

246 Visualizzazioni

See More

JoVE Logo

Riservatezza

Condizioni di utilizzo

Politiche

Ricerca

Didattica

CHI SIAMO

Copyright © 2025 MyJoVE Corporation. Tutti i diritti riservati