Introduction to Mass Spectrometry

Panoramica

Source: Laboratory of Dr. Khuloud Al-Jamal - King's College London

Mass spectrometry is an analytical chemistry technique that enables the identification of unknown compounds within a sample, the quantification of known materials, the determination of the structure, and chemical properties of different molecules.

A mass spectrometer is composed of an ionization source, an analyzer, and a detector. The process involves the ionization of chemical compounds to generate ions. When using inductively coupled plasma (ICP), samples containing elements of interest are introduced into argon plasma as aerosol droplets. The plasma dries the aerosol, dissociates the molecules, and then removes an electron from the components to be detected by the mass spectrometer. Other ionization methods such as electrospray ionization (ESI) and matrix assisted laser desorption ionization (MALDI) are used to analyze biological samples. Following the ionization procedure, ions are separated in the mass spectrometer according to their mass-to-charge ratio (m/z), and the relative abundance of each ion type is measured. Finally, the detector commonly consists in an electron multiplier where the collision of ions with a charged anode leads to a cascade of increasing number of electrons, which can be detected by an electrical circuit connected to a computer.

In this video, the procedure of ICP-MS analysis will be described by the detection of 56Fe as an example.

Procedura

1. Cleaning of Polycarbonate Tubes

  1. Use polycarbonate tubes resistant to acidic solutions for sample digestion. In order to remove any contaminating trace of iron, fill all tubes with 5 mL of 0.1 M HCl.
  2. Place tubes in a water bath for 1 h at 50 °C.
  3. Wash the tubes with 5 mL of Milli-Q water and dry the tubes in an oven or chemical hood.

2. Sample Preparation and Digestion

  1. Place 200 µL of sample in 1.8 mL of concentrated nitric acid (65

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Risultati

ICP-MS analysis of samples containing iron oxide nanoparticle is shown below. A standard curve was carried out using known concentration of 56Fe (Figure 1). The correlation coefficient being close to 1 (R2 = 0.999989) showed the good linear relationship between the sample concentrations and the intensity measured by the detector. Samples of interests showed values within the calibration range (Figure 2). The concentrations calculated by the software were then adjusted according to the d

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Tags
Mass SpectrometryAnalytical TechniqueIdentificationQuantificationUnknown CompoundsSampleStructure DeterminationGas Phase IonsMass to charge RatioQuantitative InformationQualitative InformationBasic ConceptsInstrumentationElement QuantificationIonization SourceMass AnalyzerDetectorPositive ChargeElectron Beam ImpactPlasmaLasersFragmentationMolecular Structure DeterminationHard Ionization TechniquesSoft Ionization Techniques

Vai a...

0:00

Overview

0:56

Principles of Mass Spectrometry

4:14

Sample Preparation and Digestion

5:59

Preparation of the Instrument

7:13

Instrument Operation

8:03

Applications

10:05

Summary

Video da questa raccolta:

article

Now Playing

Introduction to Mass Spectrometry

Analytical Chemistry

111.9K Visualizzazioni

article

Sample Preparation for Analytical Characterization

Analytical Chemistry

84.1K Visualizzazioni

article

Internal Standards

Analytical Chemistry

204.3K Visualizzazioni

article

Method of Standard Addition

Analytical Chemistry

319.2K Visualizzazioni

article

Calibration Curves

Analytical Chemistry

793.6K Visualizzazioni

article

Ultraviolet-Visible (UV-Vis) Spectroscopy

Analytical Chemistry

620.8K Visualizzazioni

article

Raman Spectroscopy for Chemical Analysis

Analytical Chemistry

51.0K Visualizzazioni

article

X-ray Fluorescence (XRF)

Analytical Chemistry

25.3K Visualizzazioni

article

Gas Chromatography (GC) with Flame-Ionization Detection

Analytical Chemistry

280.8K Visualizzazioni

article

High-Performance Liquid Chromatography (HPLC)

Analytical Chemistry

382.8K Visualizzazioni

article

Ion-Exchange Chromatography

Analytical Chemistry

263.6K Visualizzazioni

article

Capillary Electrophoresis (CE)

Analytical Chemistry

93.2K Visualizzazioni

article

Scanning Electron Microscopy (SEM)

Analytical Chemistry

86.8K Visualizzazioni

article

Electrochemical Measurements of Supported Catalysts Using a Potentiostat/Galvanostat

Analytical Chemistry

51.3K Visualizzazioni

article

Cyclic Voltammetry (CV)

Analytical Chemistry

124.0K Visualizzazioni

JoVE Logo

Riservatezza

Condizioni di utilizzo

Politiche

Ricerca

Didattica

CHI SIAMO

Copyright © 2025 MyJoVE Corporation. Tutti i diritti riservati