Pedicle Screw Placement Using an Augmented Reality Head-Mounted Display in a Porcine Model

The augmented reality head-mounted display, Magic Leap, was used in combination with a conventional navigation system to place pedicle screws in a porcine model by adhering to a novel workflow. With a median insertion time of <2.5 min, submillimeter technical accuracy and 100% clinical accuracy were achieved according to Gertzbein.

Rozdziały w tym wideo

0:00

Introduction

0:56

Porcine Cadaver Specimen and Identification of the Vertebral Levels of Interest

2:26

Instrument Calibration and Head‐Mounted Device (HMD) Fitting

3:43

Planning the Pedicle Screw Placement, Pedicle Placement, and Visualization of the Screw Placement

5:01

Results: Analyzing the Technical Accuracy of the Pedicle Cannulation

5:55

Conclusion